Opto. Components
Balanced Detector (Free-Space)
BALANCED DETECTOR
SwiPhotoniX Free-Space Balanced Photodetector Series is designed for high-SNR differential measurement.
Overview
Balanced Detector (Free-Space)
SwiPhotoniX Free-Space Balanced Photodetector Series is designed for high-SNR differential measurement. Precisely matched silicon (Si) and indium gallium arsenide (InGaAs) detectors are integrated with a differential amplifier circuit, achieving a common-mode rejection ratio above 40 dB to suppress laser intensity noise. The product accepts both free-space and fiber-coupled input, with the signal output through a BNC connector.
Key Features
- Wavelength Coverage 200–1700 nm
- Free-Space & Fiber Compatible
- High CMRR (>40 dB)
- Ultra-Low Noise (<3 mV)
- Proprietary Dual-Transimpedance Differential Amplification (Patented)
Specifications
Technical Specifications
| Parameter | BPDV20K | BPDV1M | BPDI20K | BPDI1M | Unit |
|---|---|---|---|---|---|
| Wavelength | 200-1100 | 200-1100 | 800-1700 | 800-1700 | nm |
| Sensor Dia. | 3.2 | 3.2 | 0.5 | 0.5 | mm |
| 3 dB Bandwidth | 20K | 1M | 20K | 1M | Hz |
| Rise Time | 17500 | 350 | 17500 | 350 | ns |
| Responsivity | 0.55@960 | 0.55@850 | 0.9@1550 | 0.9@1550 | A/W |
| Sat. Power | 1.4 | 28 | 1.4 | 28 | μW |
| Min. Power | 7 | 140 | 7 | 140 | nW |
| Output Connector | BNC | BNC | BNC | BNC | — |
| Output Impedance | 50 | 50 | 50 | 50 | Ω |
| Max Gain | 5×10⁶ | 2.5×10⁵ | 5×10⁶ | 2.5×10⁵ | V/W |
| Noise Voltage | <3 | <3 | <3 | <3 | mV |
Need help choosing a model?
Tell us wavelength range, bandwidth, signal level and environment — we reply with a configuration within 48 hours.